In-vivo color flow mapping data from carotid artery


The data were acquired with the experimental scanner RASMUS at the time of peak systole of a healthy 30 year old male. The transducer was a B-K 8812 linear array transducer.

The parameters for the measurement are given below:

Ultrasound scanner: RASMUS experimental ultrasound scanner
Scan mode:CFM and B-mode
Transducer: B-K 8812, 6.2 MHz linear array probe
Matlab file name: cfm_carotis.mat
Sampling frequency: 40 MHz
Resolution: 16 bits samples
Speed of sound c: 1540 m/s
Puls repetition frequency fprf 6 kHz
Center frequency f0 5 MHz
Focus depth: 18 mm
Cycles in pulse: 8
Start depth of data: 1 mm
End depth of data: 30 mm
Dimension left to right: x_min: -9.75 mm to x_max 9.75 mm

The data can be obtained from the ftp-server from the directory:

https://courses.healthtech.dtu.dk/22485/files/ult_data/in-vivo/cfm_carotis/

in the file cfm_carotis.mat. The size is 3.7 Mbytes.

The data can be loaded into Matlab. rf_cfm_data is the variable containing the data for velocity estimation. It is a three dimensional matrix. The first index is the sample number, the second is the emission number (1-64) and the third is the position in the image (1-16) equally spaced out between x_min and x_max. The 16 velocity lines covers the whole area scanned by the B-mode image. The data are stored as 16 bits signed integers to save space.

bmode_data contains data for the B-mode image.

D is a discriminator matrix to seperate the blood from the tissue. It has a value of 1 where the velocity should be shown in the CFM image and the value 0 where no blood is present.

CFM image from carotis

The final image made from the data should look as above.